4
Maintaining Capillary Columns
Preparing the column installation to the CFT fitting
Maintaining Your GC
41
Figure 9.
Tighten the internal nut
7
Using the 1/4-inch open-ended wrench, turn the internal nut an additional 15 to 20 degrees
clockwise to assure the ferrule is properly swaged onto the column. See
Figure 10.
Ferrule swaged to column and internal nut
Table 6
Degrees to tighten the internal nut for proper swaging
Ferrule part number
Degrees to tighten internal nut
G3188-27501
50 – 100 degrees
G3188-27502
30 – 70 degrees
G3188-27503
20 – 50 degrees
G3188-27504
60 degrees
G3188-27505
40 – 90 degrees
G3188-27506
20 – 50 degrees