7
Maintaining the COC Inlet
To Change a Septum on the COC Inlet
Maintaining Your GC
111
4
Before making an injection, check the alignment of the entire assembly using the proper
size syringe.
5
On the GC touchscreen select
Maintenance
>
Inlets
>
Septum injections
,
then touch
Reset
Counter
.
6
The GC maintenance wizard will perform checks at the appropriate times, including
Leak &
Restriction
tests, and will automatically reset the maintenance counters.
7
Select
Finished
,
then select
OK
to exit the GC maintenance wizard
.
8
Restore the analytical method.
For manual 200-µm injections with
fused silica needle