5-2
ADS Manual
Installation
A typical field installation involving external power requires the
following:
An AC mains power line terminated at a pole or post close to
the manhole in which the monitor is installed
An AC distribution device (i.e., disconnect, fuse, or breaker
box) on the pole/post
An earth ground rod installed next to the pole/post bonded to
the AC distribution device
An AC to 24-volt DC converter (i.e., power supply). ADS
recommends a power supply available through ADS (p/n
508293), which works with worldwide AC power (94 – 240
VAC, 50 – 60 Hz). This power supply is provided
automatically as part of the XIO and XBUS, but must be
ordered for the ExPAC, when a power supply is required.
However, please note that the customer may use another AC-to-
DC converter/power supply, if desired, to support the ExPAC,
provided it can supply from 9 to 36 volts of power at a
minimum of 15 watts.
A power cable running between the external DC source (i.e.,
power supply) and the ExPAC power input. The customer
must supply this cabling of the appropriate length and
specifications
A properly rated enclosure to house and protect the ExPAC and
power supply
Note:
Applications involving an XIO or XBUS do not
require an additional power cable or an enclosure because
they are included. The power is pre-run between the
power supply and ExPAC inside the XIO enclosure.
A communication cable (ADS p/n 8000-0378-xx,
where xx
represents the required length of the cable
) running between
the ExPAC/XBUS/XIO and the monitor
Summary of Contents for Intrinsically-Safe TRITON+
Page 19: ...1 12 ADS TRITON Manual Control Drawing ...
Page 25: ...1 18 ADS TRITON Manual First page of the TRITON ATEX Certificate ...
Page 27: ...1 20 ADS TRITON Manual First page of the TRITON IECEx Certificate of Conformity ...
Page 29: ...1 22 ADS TRITON Manual CSA Certificate of Compliance Page 1 ...
Page 30: ...Introduction 1 23 CSA Certificate of Compliance Page 2 ...
Page 32: ...Introduction 1 25 Sample TRITON Declaration of Conformity Page 2 ...