Maintenance and Troubleshooting 10-35
Problem
Time on the monitor clock is incorrect.
Possible
Causes
Monitor clock may be a faulty.
Monitor time zone difference for the PC clock may not
have been set correctly when the monitor was activated.
Possible
Solutions
Verify the time on the PC clock and correct if
necessary.
Reactivate the monitor to enable the clock.
Verify the time
zone
setting of the monitor. Correct and
reactivate if necessary.
Collect the data from the monitor and replace monitor if
defective.
Problem
You receive a
Device Time Out
message in
Qstart
.
Possible
Cause
Communications may be marginal.
Modem may be defective.
Possible
Solutions
Re-attempt communication with monitor. Verify signal
strength. Check a
Qstart
data collect log or check the
Status in
Qstart
. If the signal strength indicated is less
than -95 (for example: -99), consider re-orienting the
antenna and/or installing the antenna in a pedestal
above ground.
Contact your regional ADS representative.
Replace the monitor if defective.
Summary of Contents for Intrinsically-Safe TRITON+
Page 19: ...1 12 ADS TRITON Manual Control Drawing ...
Page 25: ...1 18 ADS TRITON Manual First page of the TRITON ATEX Certificate ...
Page 27: ...1 20 ADS TRITON Manual First page of the TRITON IECEx Certificate of Conformity ...
Page 29: ...1 22 ADS TRITON Manual CSA Certificate of Compliance Page 1 ...
Page 30: ...Introduction 1 23 CSA Certificate of Compliance Page 2 ...
Page 32: ...Introduction 1 25 Sample TRITON Declaration of Conformity Page 2 ...