3-58 ADS
Manual
SLIDING ADAPTER BRACKET
SENSOR MOUNTING PLATE
SURFACE COMBO/ULTRASONIC DEPTH SENSOR
PREVIOUS VERSION OF
Attaching the Surface Combo Sensor or Ultrasonic Depth Sensor to the
Sliding Adapter Bracket
12.
Confirm the sensor is still level.
To protect the crystals on the
sensor, use a block of wood of uniform dimensions between the
level and the flat face of the horizontal, ultrasonic depth sensor
portion of the sensor!
13.
Determine the physical offset for the Surface Combo Sensor or
Ultrasonic Depth Sensor by measuring the distance from the
crown (top) of the pipe to the face of the sensor (horizontal
surface with the two ultrasonic crystals on either sensor). This
value is required when configuring the monitor using the
Q
start
software.
Do not measure in reference to the the angled
surface velocity or surcharge velocity portions of the sensor
.
Summary of Contents for Intrinsically-Safe TRITON+