Configuration and Activation 6-75
2.
Allow the
monitor to log data for an appropriate
amount of time to obtain an adequate amount of data. Please
note the interval at which the monitor takes readings for
Modbus output is based on the
Scan
rate.
3.
Collect the data from the monitor. Refer to
Collecting Data
from the Monitor
on page 6-82 for detailed instructions.
4.
Compare the data collected from the monitor to the output
values obtained through Modbus. The data collected through
Qstart
reflects an average of the readings taken over a specific
interval; the values retrieved through the Modbus interface
represent the current (
persistent
) data in the monitor from the
last interval (
readings
). Therefore, the values compared may
not be exactly the same, but should be similar.