42/23-81 EN Rev. 4
System Manual SCK Gas Sampling System
201
Step Procedure
3
Unscrew adjustment screws
(1)
, turn probe until the desired posi-
tion is achieved.
Note:
To make work easier, have another person lift up the probe
tip slightly.
4
Tighten adjustment screws.
5
Unlock EMERGENCY STOP switch, turn service switch back to
OFF
position and insert probe again using the controller. Check the
position of the entire sampling pipe in relation to the mounting
flange, adjust if necessary.
WARNING!
Probe moves into the kiln.
Do not stand in the area of
travel of the probe or probe carriage until the probe is com-
pletely inserted into the kiln and the warning lamps go out.
If you have changed the position of the probe as described, you must then check
and, if necessary, adjust the end position sensor. Follow the instructions in
Probe
retractor: Checking and, if necessary, adjusting the position indicator (probe end
position)
on page 202.
After changing
horizontal position