42/23-81 EN Rev. 4
System Manual SCK Gas Sampling System
187
Gas sampling probe 60S: Replacing the seal set
twice
yearly
Always take the following safety measures when performing this work:
Wear the general protective equipment.
Turn the service switch to
ON
and secure the service switch against unauthor-
ized switching (with a padlock).
Additionally press the EMERGENCY STOP switch.
Do not enter the danger area around the retractor until it is at standstill and the
warning lamps and buzzer are off and the flap valve is closed.
Do not perform the work until the gas sampling system has cooled down.
Explanation of safety measures:
Safety regulations for maintenance work on page
156.
Wrench
SW
22
Socket
head
screwdriver
O-rings
The following conditions must be met:
The service switch is in the
ON
position, the gas sampling probe is completely
retracted from the kiln (see
Operating the service switch
on page 159).
The EMERGENCY STOP switch is pressed.
Proceed as follows:
Step Procedure
1
Allow gas sampling probe to cool down.
WARNING!
Risk of burning in the event of contact with the hot
probe.
Do not touch the probe until it has cooled down to be-
low 50 °C.
Maintenance interval
Safety
measures
Tools required
Requirements
Instructions
Summary of Contents for SCK
Page 2: ......
Page 16: ......
Page 36: ......
Page 54: ......
Page 110: ......
Page 222: ......
Page 232: ......
Page 235: ...42 23 81 EN Rev 4 System Manual SCK Gas Sampling System 235 ...
Page 236: ...236 System Manual SCK Gas Sampling System 42 23 81 EN Rev 4 Electrical connection plan ...
Page 237: ...42 23 81 EN Rev 4 System Manual SCK Gas Sampling System 237 ...
Page 238: ...238 System Manual SCK Gas Sampling System 42 23 81 EN Rev 4 Transformer connection plan ...
Page 242: ......
Page 243: ......