5 Calibration, M2000
5.4.1 Post calibration procedure
3HAC021111-001 Revision: B
129
5.4: After calibration
5.4.1 Post calibration procedure
General
Perform the following procedure after calibrating any manipulator axes. The procedure is
intended to verify that all calibration positions are correct.
Procedure
Step
Action
Illustration
1.
Run the calibration home position program twice.
Do not change the position of the manipulator
axes after running the program!
.
2.
Check the calibration positions.
If the axes are outside the tolerance, start the cal-
ibration procedure from the beginning, in increas-
ing sequence of the axes.
Detailed in section Calibration
on page 110.
3.
Repeat the check as above.
4.
Adjust the calibration marks when the calibration
is done.
Shown in section Calibration
scales and correct axis posi-
tion on page 102.
5.
The system parameters will be saved to the stor-
age memory at power off.
6.
Change the values on a new label and stick it on
top of the label located
•
underneath the flange plate on the base.
7.
Remove any calibration equipment from the
manipulator.
Summary of Contents for IRB 1400
Page 1: ...Product manual Articulated robot IRB 1400 M2004 ...
Page 2: ......
Page 3: ...Product manual 3HAC021111 001 Revision B Articulated robot IRB 1400 M2000 M2004 ...
Page 12: ...0 0 2 Product documentation M2000 10 3HAC021111 001 Revision B ...
Page 132: ...5 Calibration M2000 5 4 1 Post calibration procedure 130 3HAC021111 001 Revision B ...
Page 142: ......
Page 148: ...7 Alternative calibration 7 0 5 Retrieving offset values 146 3HAC021111 001 Revision B ...
Page 170: ...10 Foldouts 10 0 1 Introduction 168 3HAC021111 001 Revision B ...
Page 171: ......
Page 172: ......
Page 173: ......
Page 174: ......
Page 175: ......
Page 176: ......
Page 177: ......
Page 178: ......
Page 179: ......
Page 180: ......
Page 182: ...11 Circuit diagram 11 0 1 Introduction 170 3HAC021111 001 Revision B ...
Page 196: ...Index 184 ...
Page 197: ......