24
02/2010
You can change the settings of this panel using the following function elements:
Activation / deactivation of the excitation wavelengths (check box) and setting of
excitation intensities (slider). Open the
Laser Control
tool via the
Laser
icon.
Selection of the main dichroic beam splitter (MBS) or secondary dichroic beam splitter
(SBS) position through selection from the relevant list box.
Selection of a block filter through selection from the relevant list box.
Selection of an emission filter through selection from the relevant list box.
Activation / deactivation (via check box) of the selected channel.
•
Set the Pinhole diameter (via slider or input box).
•
Press the
Count rate
button to open the Real Time
display window for the detector
Count rate
in all active
channels. Adjust the Laser power In the
Laser Control
panel to obtain a satisfactory count rate.
•
Press
Adjust Pinhole
to align the pinhole for each newly
defined beam path. After adjusting the sample carrier,
align the pinhole automatically in
X
and
Y
by first
conducting a coarse and then a fine alignment.
Use the
Models
tool to define model equations to which
measured data can be fitted.
You have three options to define a model:
−
Correlation
: assemble a correlation model from
predefined equations, which will be fitted analytically
−
PCH
: assemble a photon counting histogram model,
which will be fitted numerically
−
Formula
: program a user defined model equation,
which can be fitted analytically
Fig. 32
Pinhole adjustment
Fig. 33
Models tool