LEFT TOOL AREA AND HARDWARE CONTROL TOOLS
LSM 710 and LSM 780
Carl Zeiss
Acquisition Tab
Systems
72 M60-1-0025
e
02/2010
After acquisition, the grey dynamics of the image are automatically compressed (see description above) to
match the range of the output display. This step visibly intensifies weak signals (Fig. 62).
Fig. 62
HDR image display using Sampling mode
The
Illumination
mode still provides a linear grey stepping which allows quantitative measurements.