LEFT TOOL AREA AND HARDWARE CONTROL TOOLS
LSM 710 and LSM 780
Carl Zeiss
Acquisition Tab
Systems
78 M60-1-0025
e
02/2010
If the defined free shape curve becomes too complicated or the selected
Scan Speed
is too high, the
following message appears in the status bar:
Maximum scanner acceleration exceeded!
•
In this case, reduce the
Scan Speed
set in the
Acquisition mode
tool.
•
If the generated contour and the line traced by the scanner are not in coincidence, reduce the
Scan
Speed
by a further amount.
If no sufficient coincidence of the two lines can be achieved by the reduction of the scan speed, you have
to calibrate the scanner position signal.
5.2.6.4
Spot Mode
In the
Spot
mode
fluorescent or reflected light
occurring from a single voxel xyz is detected. In this
mode a spot can be defined by two perpendicular
lines in the
Image
using the
Spot Select
button
which is located in the lower part of the
Acquisition Mode
tool window (Fig. 70).
The following parameters for scanning a spot can
be defined:
Objective, Speed, Pixel Depth and Offset in X
and Y (= position within the image)
These functions operate as described in section
Frame Mode for LSM
.
In
Spot
mode it is not possible to acquire a
Z-Stack
. After definition of the spot position the
only possible scan mode is
Time Series
.
Fig. 70
Acquisition Mode tool - Spot Mode