LEFT TOOL AREA AND HARDWARE CONTROL TOOLS
LSM 710 and LSM 780
Carl Zeiss
Acquisition Tab
Systems
102 M60-1-0025
e
02/2010
•
The
Imaging Setup
tool now shows the active
channel and the detection range for the signal.
You may want to save these settings as a
specific imaging configuration (Fig. 99).
5.2.19.2
Defining the Scan Parameters
•
Open the
Acquisition Mode
tool (Fig. 100).
•
Select
Frame
mode and a
Frame Size
of 512 x
512 to start with.
•
The
Optimal
button automatically adjusts the
frame size to match the magnification and
resolution of the objective. Thus, images are
acquired without collecting redundant image
information.
•
The Scan Speed is set to 9 per default. A lower
speed produces better images as the light
collection per pixel is higher. Use a speed from
7 to 9 to begin with.
•
Start with the following settings on the
Averaging
panel:
Bit depth:
8 bits
Number:
1 (no averaging)
Additional parameters like scan direction are displayed in
Show all
mode. Scan direction is per
default set to unidirectional.
The default settings for Acquisition Mode are suitable for good image quality and do not need
to be changed for initial scanning.
Using the
Live
button is a convenient way of creating an overview scan.
Fig. 99
Imaging Setup tool
Fig. 100
Acquisition Mode tool