LSM 710 and LSM 780
LEFT TOOL AREA AND HARDWARE CONTROL TOOLS
Systems
Acquisition Tab
Carl Zeiss
02/2010 M60-1-0025
e
75
Scan Area
In this panel, the scan field is set for zoom, rotation
and offset in relation to the field of view of the
microscope. The diagonals of the outer square on
the left-hand side correspond to the field of view
of the microscope.
The inner square contained in it (rectangle in the
case of differently set frame size) represents the
scan range and immediately shows the changes
made to zoom, rotation and offset.
•
Set the desired
Zoom
factor via the slider or by clicking on the small red squares outlining the inner
square.
−
The
Zoom
factor can be set continuously in the range from 0.6 to the maximum of 40, and is
displayed in the relevant input box. The value 0.6 corresponds to factor 1, and value 40 to factor
66, related to the field of view. From zoom factor 5 onwards, the magnification will be empty.
−
Recommended setting to start with: Zoom 1.
•
To rotate the scan area, use the slider or click on the blue dot in one of the corners of the inner
square.
−
Recommended setting to start with: Rotation 0°.
•
Move the scan area by clicking on
for offset in horizontal direction
and
for offset in vertical direction.
•
The offset of the scan area from the center of the field of view is displayed online in µm for X and Y.
−
Clicking, holding and drawing the rectangle with the mouse permits the scan area to be moved
directly within the field of view.
−
Recommended setting to start with: Offset X = 0, Y = 0
During the scan procedure, the functions
Speed
,
Scan Corr
,
Zoom
,
Rotation
and
Offset
can
be influenced online.
By clicking on the
Reset All
button the scan zoom is set to 1, the XY offsets are set to the zero position
and the ratio angle is set to 0°.
Fig. 67
Scan Area panel