Running the Calibration Matrix
If you haven’t selected the check box for Basic Calibration, the
Calibration Matrix
will be
selected automatically. In this case, when the Calibration Matrix is running, the Basic Calibration
is disabled.
•
This window provides three check boxes
(Fig. 14)
•
The recommended setting is to select
Create
new calibration file.
All previous files will be
deleted.
Exceptions:
No check box checked
−
Especially after an unexpected interruption
of the program, it is possible to continue
from the position where the process stopped
(saving time).
Overwrite Collimator values
−
The values for the chosen objective will be
added or updated.
−
After selecting the collimator option, the
Pinhole Y option will be selected
automatically.
•
Double check if the objective settings (shown at
the bottom of the macro window) match the
currently used objective.
•
Click the
Calibrate
button.
Meaning of the Graph (Figure):
Brown:
Focal plane of the CCD Line Detector
Red:
Focal plane of the collimator
Orange:
Estimated focal plane
The brown and red graphs should be close to this
curve.
Fig. 14
Running the Calibration Matrix
Fig. 15
Graph of Focus Function