TOOLS, ADDITIONAL SOFTWARE
LSM 710 and LSM 780
Carl Zeiss
3D for LSM
Systems
34 M60-1-0025
e
02/2010
If
Blue/Red
is selected, the voxels with grey values within the interval
L
ow,
H
igh remain unchanged.
Voxels with grey values less than
L
ow are highlighted in blue; those with grey values higher than
H
igh
are highlighted in red.
If the
Invert
option is selected, the grey values outside the defined interval will be segmented.
If the option
Binary
is selected, then all grey values in the range from
L
ow to
H
igh will be set to white
(grey value 255) in the
Output
image sequence, while all others will be set to black (grey value 0). If the
option is not selected, the grey values within the selected interval remain unchanged, while those outside
the range will be set to black. The measurement function accepts both results without any difference in
the results.
Parameters:
Input
Input image sequence
Output
Resulting image sequence
Colour
Green - Selected interval is displayed in green
Blue
/
Red
Grey values below the selected interval are displayed in blue, grey values
above in red
Binary
0 - Selected voxels retain the original grey value
1 - Selected voxels are set to grey value 255, the rest to grey value 0
Invert
0 - Grey values inside the selected interval are segmented
1 - Grey values outside the selected interval are segmented
L
Low grey value threshold
C
Center of threshold interval
H
High grey value threshold