CONFOCOR
3
ConfoCor 3
Action Buttons
Carl Zeiss
02/2010 M60-1-0025
e
111
9.12.3
XYZ-Scan
•
Press
xyz-Scan
to start a Scan as defined in the
X-Y-Z Scan tool
.
•
This opens the
Count Rate Scan
window.
•
The focus is now moved either in the x, y, or z
direction or a combination of two (xy, xz and yz)
in dependence of what button is active over the
pre-selected distance. During the scan the count
rate is displayed for the selected channels.
Dependent on the scan modalities the count
rates for the selected channels are plotted in a
line plot (x, y and z) (Fig. 114) or frame plot (xy,
xz and yz) (Fig.
115). In the frame plot
intensities are color coded. The maximum
intensity in Hz is displayed in the
Maximum
count rate channel 1
and
2
display boxes.
Please note, that x and y scans are
performed by the stage regardless if the
Scanner button is pressed or not in the
Measurement
tool group. Z scans are per-
formed by the nosepiece.
Scans will be performed relative to the actual
position. The actual position will always be
displayed as the 0 coordinates.
A maximum (or minimum) in the resulting
curve or plot usually indicates a region of
interest.
•
When the scan has been finished, the vertical
black line (line plot) or the cross-lines (frame
plot) can be used to re-position the focus. Just
drag the bar or cross-lines with the left mouse
button held down to the new position and
release the mouse. By clicking the
OK
button
the stage or nosepiece will move to the new
position and the Scan window closes. The
Abort
button will be inactivated for a finished
scan. You can press
Ok
or
Abort
during the
scan to interrupt the scan. Ok will close the
window and the old position is kept. Abort
does not close the window, so you are still able
to reposition. Press
Ok
to leave the Scan
window.
Fig. 114
Z-Scan result window (line plot)
Fig. 115
XY-Scan result window (line plot)