LSM 710 and LSM 780
NOTES ON DEVICE SAFETY
Systems
Environmental Requirements
Carl Zeiss
02/2010 M60-1-0025
e
9
1.5
Environmental Requirements
1. Operation, specified performance
T = 22 °C ± 3 °C without interruption (24 h a day
independently whether system is operated or
switched-off)
2. Operation, reduced performance
T = 15 °C to 35 °C, any conditions different from
item 1. and 5.
3. Storage, less than 16 h
T = -20 °C to 55 °C
4. Storage, less than 6 h
T = -20 °C to 55 °C
5. Temperature gradient
± 0.5 °C/h
6. Warm up time
1 h, for high-precision and/or long-term measure-
ments
≥
3 h
7. Relative humidity
< 65 % at 30 °C
8. Operation altitude
max. 2000 m
9. Loss of heat
4 kW
These requirements do not include the requirements for high precision measurements. Please refer
to the Operator’s Manual of the microscope for these requirements.
1.6
Notes on Setting up the Microscope System
Installation and commissioning of the LSM 710 and LSM 780 systems and the ConfoCor 3 systems
must be performed by authorized Carl Zeiss service staff. The system should not be used prior to
instruction by a Carl Zeiss representative.
The LSM 710, LSM 780, LSM 710 NLO and LSM 780 NLO laser scanning microscopes and the
ConfoCor 3 spectroscope are delivered in several crates.
The LSM 710 and LSM 780 systems and ConfoCor 3 must be set up so as to ensure that the
minimum clearance between the wall and the rear of the system is no less than 0.5 m. This
clearance is needed for adjustment and maintenance operations.
Do not set up the unit in the proximity of heat sources such as radiators or direct sunlight. To avoid heat
build-ups, the ventilation slots on the microscope system must not be covered up.
The system must not be set up in areas with potential danger by explosives.
The unit must be connected to a properly installed socket outlet with earthing contact by means of the
mains cables supplied. Continuity of PE connection must not be affected by the use of extension leads.