LSM 710 and LSM 780
CENTER SCREEN AREA / IMAGE CONTAINERS - …
Systems
Unmixing View
Carl Zeiss
02/2010 M60-1-0025
e
245
Tools in the Unmixing - Extract View Options Control Block:
These tools (Unmixing – Extract) are only available for Lambda stacks on an LSM 710 and LSM 780
system with 34 channel Quasar detector. Hence, they are not available for LSM 700 or LSM 7
LIVE
Hardware (tool tab grayed).
Button:
Slider:
Selection of a wavelength range (use the sliders) that will be assigned to
a channel. This assignment can be used for scanning (when
Apply to
Hardware
is pressed, see below) and /or for extracting a single or
multichannel image from the Lambda stack (when
Extract Channels
is
pressed). Clicking on the color box in the channel button opens the
color selection list and a LUT can be assigned to this channel.
Add (+) or reduce (-) the number of channels displayed and available for
assignment. Up to 8 channels labeled after the assigned LUT are
available.
Sum signals from selected detector elements
Mean signals from selected detector elements
Autoscaling of output channels (individually)
channels
button
Generates a new image or multicolor images based on the settings
made in the channels. Two or more Lambda channels are binned to
form the channels 1 ... 8. The generated image is displayed in a new
Image Display
window and is no longer a Lambda stack.
button
Sets the QUASAR detector parameters to match the previously defined
spectral ranges / channels. For the used in binning mode as specified in
the settings ChS1, ChS2, ChS3 ... ChS8. See
Single Track
and
Multi
Track
in
Config Control
for the effect of the action. The
Lambda
stack
setting is not affected by this action.