10-6
IM DL350-01EN
1-Cycle Mode (1-Cycle Mode)
Instead of automatically measuring the measurement time period specified by Time Range1 and Time Range2,
you can automatically measure the first period after Time Range1.
The method of determining the period is the same as the method for determining the Period measurement item.
In this mode, after the period is determined, the values of the measurement items related to voltage and area
are computed. This mode is effective for measurement items, such as Rms or Avg, that may result in errors
depending on the measurement time period setting.
The measurements of time axis items and X-Y areas are not affected.
• OFF: 1-cycle mode is disabled.
• ON: 1-cycle mode is enabled.
Time Range1
Time Range2
Distal line (90%)
Proximal line (10%)
Mesial line (50%)
High(100%)
Low(0%)
1 Cycle
OFF
Measurement Time Period for Each Cycle Mode
If the space between Time Range1 and Time Range2 is less than one period, “*****” is displayed for the
measured value.
Notes about Automated Measurement of Waveform Parameters
• Measurements cannot be made on a snapshot waveform or an accumulated waveform that is not the most
recent waveform.
• When an item is impossible to measure its measured value is displayed as ‘*****’.
• When the measurement time period exceeds 100 Mpoint on a SD recording waveform
• When the delay measurement mode is Degree and the trigger source is set to the reference of the
reference waveform
• The instrument may not measure correctly if the waveform amplitude is small.
• If there are two or more waveform periods within the measurement time period, the time-axis parameters are
measured on the first period.
• When automated measurement is performed on an SD recorded waveform, extra time will be required for
processing.
• Automated measurement may require additional time depending on settings such as the memory length, the
number of measurement items, and the input waveform. During automated measurement,
appears in
the center of the screen.
• To cancel automated measurement, set Mode to OFF. Measurement stops immediately.
• For FFT waveforms, only the Max and Min items can be measured. For a power spectrum, only the overall
value can be measured.
10 Automated Measurement of Waveform Parameters