![YOKOGAWA ScopeCorder DL350 Скачать руководство пользователя страница 25](http://html.mh-extra.com/html/yokogawa/scopecorder-dl350/scopecorder-dl350_user-manual_911346025.webp)
4-3
IM DL350-01EN
Linear Scaling (Linear Scale)
You can configure the linear scaling settings of all channels while viewing the settings in a list. This function is
the same as the linear scaling function for voltage measurement.
Linear scaling (Linear Scale); A and B for AX+B or P1:X, P1:Y, P2:X, and P2:Y for P1-P2; unit (Unit); display
mode (Disp Type); number of decimal places (Decim Num); unit prefix (Sub Unit)
The 701271 (STRAIN_DSUB) strain module supports
Operation (Operation)
Copy (Channel Copy)
You can copy the various vertical axis and linear scaling settings from one channel to other channels whose
modules are of the same type as the source channel.
• Source Channel (Source Channel)
Set the source channel to a channel from CH1 to CH6 or to 16chVOLT or 16chTEMP/VOLT.
*
* When a 16-CH voltage input module or 16-CH temperature/voltage input module is installed. After you
select 16chVOLT or 16chTEMP/VOLT, select a sub channel.
• Destination Channel
Set the destination channel to a channel from CH1 to CH6.
• Execute (Execute)
Select Execute to copy the settings.
• When a 16-CH Voltage Input Module or 16-CH Temperature/Voltage Input Module is installed, you can also
copy the settings of the sub channels. You can specify only channels as copy destinations.
• The waveform display color and label settings are not copied.
Strain Balance and DC Offset Cancellation(Balance & DC Offset Cancel)
Cancel the DC offset on multiple installed strain modules.
If a strain module is installed, stain balancing is executed on the strain module.
• CH1 to CH4
Select the modules that you want to perform strain balancing and DC offset cancellation on.
• Execute (Execute)
Execute strain balancing and DC offset cancellation.
For information about strain balancing, see the section on strain measurement.
For information about DC offset cancellation, see the section on DC offset cancellation.
You cannot select channels in which strain modules are not installed.
4 Vertical Axis