10-13
IM DL350-01EN
Detail Settings (Detail Setup)
Set the reference level that is used to measure various parameter values, such as the rise and fall times, for
each measurement source waveform.
Source Waveform (Trace)
Select the trace for setting the details.
CH1 to CH6
1
, 16chVOLT
2
, 16chTEMP/VOLT
2
, CAN
3
, LIN
3
, SENT
3
, Math1 to Math2
*1 You can select the channel of an installed module. On a 4-CH module, select sub channel 1 or 2. You cannot
select the channel of a logic module.
*2 When a 16-CH voltage input module or 16-CH temperature/voltage input module is installed. After you select
16chVOLT or 16chTEMP/VOLT, select a sub channel.
*3 On a model with the /VE option when a CAN bus monitor, CAN & LIN bus, or SENT monitor module is
installed.. After you select CAN, LIN, or SENT, select a sub channel. This cannot be selected on a CAN bus
monitor or CAN & LIN bus monitor module if the data type (Value Type) is set to Logic. On a SENT monitor
module, S&C and Error Trigger sub channels cannot be selected.
• Distal, Mesial, and Proximal Settings (Mode)
Set the method for setting the three levels that are used as references in the measurement of various
parameter values, such as the rise and fall times.
• %
You can set the distal, mesial, and proximal values as percentages of the specified trace. The high value of
the specified trace is equal to 100.0%, and the low value is equal to 0.0%.
• Unit
You can set the distal, mesial, and proximal values of the specified trace by specifying physical values,
such as voltages or temperatures.
• Selecting What You Will Set (Target): Distal, Mesial, or Proximal
Select which level you will set: Distal, Mesial, or Proximal.
• Distal, Mesial, and Proximal Settings (Distal, Mesial, Proximal)
You can set the distal, mesial, and proximal values.
Selectable range: 0.0 to 100.0% (in 0.1% steps) or voltage or temperature values that correspond to ±10
div (the steps that you can set voltage and temperature values in vary depending on the module. See
• High/Low Specification Method (High/Low)
The high and low levels are the 100% and 0% levels used to measure various parameter values, such as the
rise and fall times. You can choose one of the following methods for setting the high and low levels.
• Auto
The instrument sets the high value to the high amplitude level and the low value to the low amplitude level
based on the voltage level frequency of the waveform in the measurement time period while taking into
account the effects of ringing, spikes, etc. This method is suitable for measuring square waves and pulse
waves.
• Max-Min
The instrument sets the high and low values to the maximum and minimum values in the measurement
time period. This method is suitable for measuring sinusoidal and saw waves. It is not suitable for
waveforms that have ringing and spikes.
Saving Automated Measurement Values of Waveform Parameters (Save)
Notes about Using the 16-CH Voltage Input Module (720220)
Notes about Using the 16-CH Temperature/Voltage Input Module (720221)
10 Automated Measurement of Waveform Parameters