Fundamental Frequency (Frequency)
Set the fundamental frequency.
50 Hz, 60 Hz, Auto
Start Point (Start Point)
Set the start point of harmonic analysis.
Selectable range: –5 div to +5 div
• Scope mode: Set the positions in divisions.
• Recorder mode: Set the positions in terms of the time (ddhhmmss) corresponding to divisions.
The start and end points are similar to the settable range of the cursor display position in cursor measurement.
For details, see “Selectable Range of Cursor Positions.”
Harmonic Analysis on Voltage and Current (Line RMS Setup)
Performs harmonic analysis on voltage and current.
Analysis Mode (Mode)
• ON: Harmonic analysis is performed on voltage and current.
• OFF: Harmonic analysis is not performed on voltage and current.
Math Source Waveforms (Source)
The modules described in “
Modules on Which Harmonic Analysis Can Be Performed
” under “Harmonic Analysis
(Harmonic)” are applicable.
Hysteresis (Hysteresis)
.
Harmonic Analysis on Active Power (Power Setup)
Performs harmonic analysis on active power.
Analysis Mode (Mode)
• ON: Harmonic analysis is performed on active power.
• OFF: Harmonic analysis is not performed on active power.
Wiring System (Wiring)
The following three wiring systems are available on the instrument.
: 1P2W
: Single-phase three-wire
: Three-phase three-wire
To apply voltage, use a passive probe.
For details on how to select the appropriate passive probes and how to connect them (high and low), see
section 2.10 in the Getting Started Guide, IM DL350-03EN.
To apply current, use a current probe.
For details on how to select the appropriate current probes and how to connect them (current direction), see
section 2.10 in the Getting Started Guide, IM DL350-03EN, and the user’s manual that came with the current
probe.
15 Harmonic Analysis