![YOKOGAWA ScopeCorder DL350 Скачать руководство пользователя страница 165](http://html.mh-extra.com/html/yokogawa/scopecorder-dl350/scopecorder-dl350_user-manual_911346165.webp)
10-11
IM DL350-01EN
Result Display (Result Window)
Displays a list of calculated statistics. You can display the list of statistics when Mode is set to Cycle Statistics or
History Statistics.
Numbers are assigned to the data in order from the oldest cycle data or history data, and the automated
measurement results for each number are displayed.
The maximum and minimum values for each parameter are indicated on the list by (maximum value) and
(minimum value). If the same value appears in multiple locations, the oldest occurrence of the value is marked
as the maximum or minimum value.
The number of data points that can be listed is 64000. If the number of data points exceeds 64000, the most
recent data points in the history waveform or automatically measured item data are displayed. If the number of
data points exceeds 64000, the maximum and minimum values may be outside of range of the displayed list.
When this happens, (maximum value) and (minimum value) will not appear.
In cyclic statistical processing, you can select a waveform by tapping a waveform on the result display list to
display one period of the waveform with the selected number on the zoom display.
10 Automated Measurement of Waveform Parameters