5-16
IM DL350-01EN
Save a Screen Capture (Save Image)
The instrument saves the screen capture data to the specified destination (SD card, USB storage device, or
network drive).
• File Path (File Path), Auto Naming (Auto Naming), File Name (File Name)
Changing the auto-naming and file-name settings for saving image data will change the corresponding
settings under Image Save in the Save/Load menu.
Send Email (Mail)
The instrument sends an email to the specified address. Set the email address by pressing Utility and then
selecting Network > Mail.
• Email Send Count (Mail Count)
This is a setting in scope mode.
Set the number of email transmissions.
• Infinite: Email continues to be transmitted until you stop the action.
• 1 to 1000: Email transmission stops when the number of sent emails reaches the specified count.
In recorder mode, the number of email transmissions is 1. As such, the above settings are not displayed.
Notes about Action
• You cannot change settings while the action feature is active.
• The actions may be slow if there is network access while the following operations are being performed.
Printing and saving of screen capture data and saving of waveform data
• When SD recording is enabled, the action feature cannot be used.
Notes about the “Save Data” and “Save Image” Actions
• Do not set the storage medium’s root folder as the save destination. A file whose name is longer than eight
characters will be counted as two files. If such files exist, the number of files that can be stored will decrease.
• The maximum number of files that can be created in a single folder is 1000. Make sure that there are no files
in the destination folder before you start the action feature.
• If you select save waveform data (Save Waveform) and save a screen capture (Save Image) at the same
time, specify separate folders to save to.
• If you set Auto Naming to Numbering, as the number of saved files increases, the amount of time required to
save a file will also increase.
• Save Destination Folder during Action Execution
In the specified drive, a folder is automatically created with the date (year, month, and day) as its name, and
data is saved to that folder using file names specified by the auto naming feature. If the number of files in the
save destination folder exceeds 1000, a new folder is automatically created with the date and an incremented
sequence number (000 to 999) as its name, and the data continues to be saved in the new folder.
You can configure the instrument so that data is saved to the specified folder, not to the folder that is
automatically created with the date.