Movement Target Marker (Marker No.)
Select the marker you want to set the position of.
Position (Position)
Set the position of the selected marker.
You can move the markers within the range of −5 to 5 div of the frequency axis in 0.01 div steps.
Peak Cursors (Peak)
In the frequency range defined by FFT1 Range1 and Range2 and the frequency range defined by FFT2 Range1
and Range 2, the instrument detects peaks (Peak1 and Peak2) and measures their frequencies and levels. You
can set the two frequency ranges in the range of −5 to 5 div.
Peak Settings (Peak Setup)
The following values at the peaks are measured.
F1
Frequency at Peak1
F2
Frequency at Peak2
Y1
Level at Peak1
Y2
Level at Peak2
Range Source Waveform (Range)
Set the peak measurement source waveform to one of the waveforms below.
• FFT1: The waveform in the FFT1 window is measured.
• FFT2: The waveform in the FFT2 window is measured.
Upper and Lower Limits of the Range (Range 1/Range 2)
Set the range for detecting peaks on the frequency axis in the range of –5 to +5 div.
Linking the Upper and Lower Limits of the Range
You can change the upper and lower limits while keeping the interval between them constant.
Notes about Cursor Measurement
FFT Waveform Saving (Save)