• When you measure strain, be sure to perform balancing.
• Set the bridge voltage to 5 V or 10 V only when the bridge resistance is 350 Ω or greater. The strain will
not be measured properly if you apply a bridge voltage of 5 V or 10 V when the bridge resistance is 350 Ω
or less.
• When you use a strain gauge transducer, be sure to use it at a bridge voltage that is within the transducer’s
recommended voltage range.
• Balancing cannot be performed if there is no strain measurement bridge (bridgehead) or strain gauge
transducer connected to the channel that you are trying to perform balancing on.
• When you perform balancing on all channels, if balancing fails on even one of the channels, an error
message will appear along with information about the channels for which balancing failed.
• After you turn on the power, connect a new strain gauge, or change the measurement range, bridge
voltage, or gauge factor, you need to perform balancing again before you perform measurement.
The scale value is displayed as “imbalance” immediately after you turn on the power or when you change
the range. In this kind of situation, perform balancing.
• When you change the unit of a channel, the units for all the items related to the channel change. This
includes the upper and lower limits (Upper/Lower), trigger level (Level), automated measurement values of
waveform parameters, and cursor-measurement values.