4-63
IM DL350-01EN
If Slow CH message format:
Enhanced (ID 4bit + Data 12bit)
, Start Bit: 0, Bit Size: 16
Bit number
0
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
Bit Size: 16
0
1
1
1
1
1
1
0
1
0
ID (7-4)
Data Field (15-12)
6 bit CRC
Data Field (11-0)
bit #2
bit #3
Value Type
Select the type of data to extract.
Unsigned: Unsigned integer
Signed: Signed integer
SENT Data Conversion Conditions
Set the conversion conditions for data type Fast CH and SlowCH.
Factor/Offset
The Factor and Offset values that you set here are used to convert the extracted data to physical values.
Factor: Scaling coefficient (value per bit)
Offset: Offset value
Selectable range: –10.000E+30 to +10.000E+30
Unit
Using up to 16 characters, enter the unit to display for the waveform.
Error Count Reset (Error Count Reset)
You can reset the error count to zero.
SENT Data Display Settings (Display)
Configure the display label, zoom magnification, scaling method, and display range settings of the specified sub
channel.
For the association of each sub channel and the data type, see “
SENT Data Acquisition Destination
Zooming by Setting a Magnification (V Zoom)
Configure this setting when the data type is set to S&C (Status & Communication) and Error Trigger.
Selectable range: x0.1 to x3.33
Display Range (Upper/Lower)
Set the upper and lower limits of the waveform display range. Configure this setting when the data type is set to
Fast CH, SlowCH, or Error Count.
Selectable range:
−
30.000E+30 to +30.000E+30
4 Vertical Axis