10-5
IM DL350-01EN
• The voltage level of the detected point is the mesial line.
• The measurement item name that appears when the measured values are displayed is (Delay).
• If Mode is set to Degree and Reference is set to Trigger, the measured value is displayed as ‘*****’.
• If you set the delay measurement's Reference to Trace, measurements will not be performed when the
sample rates of the base waveform and the measurement source waveform are different. The measured
value is displayed as ‘*****’.
Measurement Time Period (Time Range1/Time Range2)
Set the measurement time period using two vertical cursors. The position of the thin dashed line (Time Range1)
is the measurement start point. The position of the thick dashed line (Time Range2) is the measurement end
point. However, for SD recording, the maximum number of data points that are measured is 100 Mpoint from
Time Range 1.
Selectable range: –5 div to +5 div
• Scope mode: Set the positions in divisions.
• Recorder mode: Set the positions in terms of the time (ddhhmmss) corresponding to divisions.
The start and end points are similar to the settable range of the cursor display position in cursor measurement.
For details, see “Selectable Range of Cursor Positions.”
10 Automated Measurement of Waveform Parameters