6-2
IM DL350-01EN
Signal Type and Trigger Type Combinations
The signal type (analog or logic) determines what trigger types you can use.
Scope Mode
CH1 to CH6, GPS
Mixed
Analog Signal
(including sub channels)
Logic Signal
Bit 1 to Bit 8
Simple
Yes
Yes
−
Edge On A
Yes
Yes
Yes
OR
Yes
Yes
Yes
AND
Yes
Yes
Yes
Period
Yes
Yes
Yes
Pulse Width
Yes
Yes
Yes
About the Wave Window trigger
Recorder Mode
CH1 to CH6, GPS
Mixed
Analog Signal
(including sub channels)
Logic Signal
Bit 1 to Bit 8
Edge
Yes
Yes
−
OR
Yes
Yes
Yes
AND
Yes
Yes
Yes
Basic Trigger Settings
•
: The trigger source signal.
•
: Specifies which edge, rising or falling, the instrument will trigger on.
•
: The trigger determination level.
•
: The trigger level margin (the instrument does not trigger on changes in the signal level
within this margin).
•
: The position where the trigger point will be displayed (applies to all trigger types).
•
: The delay from the trigger point (applies to all trigger types).
6 Triggering