7. OPERATION VIA BRAIN TERMINAL
IM 1E7B0-02E
7-10
(2) 4 to 20 mA Low Limit
Display
Description
Setting Range : -20 to 100%
Default : -20%
DATA
ESC
DIAG
PARAM
G01:4-20 LOW CUT
0 %
G02:4-20 LOW LMT
10 %
G03:4-20 H LMT
110 %
4mA
Output
5.6mA(10%)
(3) 4 to 20 mA High Limit
Display
Description
Setting range : 0 to 120%
Default : 120%
18.4mA(90%)
Output
PARAM
G01:4-20 LOW CUT
0 %
G02:4-20 LOW LMT
-20 %
G03:4-20 H LMT
90 %
DATA
ESC
DIAG
7.2.10 Forward and Reverse Flow Measurement (Refer to 6.3.8)
Display
Description
PARAM
B09:HIGH MF
1.0000
B10:OUTPUT FUNC
BI DIRECTION
B11:4-20 ALM OUT
2.4mA OR LESS
DATA
ESC
DIAG
DATA
ESC
DIAG
PULSE OUT
ALARM OUT
BI DIRECTION
AUTO 2 RANGES
LOW ALARM
TOTAL SWTICH
Pulse output
Alarm output
Forward / reverse flow measurement
Automatic two range switching
Alarm output at low flow limits
Totalization switch
PARAM
B30 :REV. SPAN
B31 :BI. DIREC HYS
B33 :FOR. SPAN2
Select "BI DIRECTION" (Direct / reverse flow measuring mode) in
parameter number "B10 ".
Reverse direction span can be set in parameter No."B30".
Flow rate unit is the same as forward direction span.
Further reverse range span should be set in the same number of places of
decimals as forward range span.
Example: forward flow rate : 1.000 then reverse flow rate should be 4.000.
Hysteresis width at switching direction can be set in parameter No."B31".
It is the rate(%) of the smaller span, either forward or reverse span.
Code
Content