6. FUNCTION AND DATA SETTINGS
IM 1E7B0-02E
6-2
(2) Span Setting using Display Keys (Example setting: 30.00 m
3
/h)
•
Span Value Setting
Switch Operation
Display
Description
SET SHIFT INC
SET SHIFT INC
SET SHIFT INC
SET SHIFT INC
SET SHIFT INC
SET SHIFT INC
SET SHIFT INC
SET SHIFT INC
SET SHIFT INC
SET SHIFT INC
SET SHIFT INC
Press the SET key during flow rate data display mode
to display setting mode shown in the left figure.
Press the SHIFT key to cause the second digit from
the left to flash.
Press the INC key to call up parameter number 03.
Press the SET key to move the flashing segment to
the first digit in the data area.
Press the INC key to set the first digit to 0.
Press the SHIFT key to move the decimal point.
Press the INC key to move the decimal point to "000.00".
*To set a "00000." as the data area is a four digit LCD,
the three digits and the space will be displayed as "000".
Press the SHIFT key to move the flashing segment two
digits to the right.
Press the INC key to enter "3".
Press the SET key once to cause all the data to flash.
*As the data display now will be scrolled, it may not
appear as shown in the left figure.
Wait 2 to 3 seconds when the data starts flashing. Then
press the SET key to return the flashing segment to the
leftmost digit.
*This display will also be scrolled to display 5-digit data.
Defalt is set as 1.0000. Change this into 030.00.
(Setting is now completed)
Note : 5-digit span data can be displayed, but only digits will be supplyed at one time. As a result, when the last
digit is changed, the last digit and the space will be displayed as the last three digits. (See 5.2.2 "Display
of 5-digit data".)