![YOKOGAWA AE100D Скачать руководство пользователя страница 126](http://html.mh-extra.com/html/yokogawa/ae100d/ae100d_instruction-manual_911083126.webp)
11. PARAMETER SUMMARY
IM 1E7B0-02E
11-3
No.
AE
BT
AE
BT
Name
Data Range, Units
Deci-
mal
Point
Initial
Setting
R/W
Description
SET
TAG NO.
DAMPING
FLOW SPAN
FLOW UNIT
TIME UNIT
SIZE UNIT
NOMINAL SIZE
LOW MF
HIGH MF
OUTPUT FUNC
4 - 20 ALM OUT
8 ASC
II
characters
0.1 to 200.0
1 to 30000
km3(10
3
X
m
3
)
m3
l (liter)
cm3(10
-2
X
m)
3
Mgal
kgal
gal
mgal
kbbl
bbl
mbbl
ubbl
m
ft
/d
/h
/m
/s
mm
inch
1 to 3000.0
0.2500 to 3.0000
0.2500 to 3.0000
PULSE OUT
ALARM OUT
BI DIRECTION
AUTO 2 RANGES
LOW ALARM
TOTAL SWITCH
2.4mA OR LESS
4.0mA
HOLD
21.6mA OR MORE
Major parameter setting items
Specifies tag number using up to
8 characters.
Sets output time constants.
Flow rate span
A combination of B04 and B05
units are used.
Select volume units of flow rate
span.
Select time constants of flow rate
span.
Sets unit of meter tube size
Sets meter tube size using B06
units.
Select low frequency
measurement meter factor.
Select high frequency
measurement meter factor.
Select pulse output / alarm output
or status output functions.
Select current to be output
during alarm. However, a
hardware failure that causes a
burn-out will result in 0 mA
output.
B00
B01
B02
B03
B04
B05
B06
B07
B08
B09
B10
B11
S
S
00
01
02
03
04
05
06
07
08
09
10
11
12
13
00
01
02
03
00
01
S
00
01
02
03
04
05
00
01
02
03
02
03
04
05
06
07
08
09
10
11
1
0 to 4
1
4
4
R
W
W
W
W
W
W
W
W
W
W
W
(Space)
3.0
1.0000
m
/s
mm
(Set)
(Set)
(Set)
PULSE
OUT
2.4mA
OR LESS
Содержание AE100D
Страница 2: ......
Страница 333: ...Blank Page ...
Страница 395: ...Blank Page ...
Страница 401: ...Blank Page ...