![YOKOGAWA AE100D Скачать руководство пользователя страница 181](http://html.mh-extra.com/html/yokogawa/ae100d/ae100d_instruction-manual_911083181.webp)
IM 1F6A0-01E
3-10
3. INSTALLATION
3.7.2 Indicator Removal and Rotation
If necessary for servicing of amplifier, remove the indicator
as follows.
(1) Turn the power OFF.
(2) Remove the cover.
(3) For the indicator, disconnect the cable connector from
the amplifier unit.
(4) Loosen the two indicator mounting screws using a
Phillips screwdriver.
(5) Pull out the indicator.
(6) Reinstall the indicator in the reverse order to its removal
(above) and secure the mounting screws.
F030702.EPS
Indicator
90
°
Indicator Mounting Screws
(2PCS)
Figure 3.1
Removing and Reinstalling the Indicator
3.7.3
Amplifier Unit Removal and Assembling
The amplifier unit can be removed as follows.
(1) Turn the power OFF.
(2) Remove the converter cover.
(3) Remove the indicator totalizer according to the proce-
dures described in paragraph 3.2.
(4) Loosen the terminal screws and remove the amplifier
unit.
The amplifier unit can be assembled as follows.
(1) Put two-mounting pins
1
into mounting hole
2
.
(2) Push the head of two-mounting
1
head lightly.
(3) Push head of IC
5
and mount the amplifier unit
3
.
(4) Tighten a mounting screw
4
.
IMPORTANT
Do not turn the amplifier unit for removal or mounting.
The connector pins may be damaged.
F030703.EPS
Amplifier Unit
Mounting Screw
Mounting Holes
Mounting Pins
IC
1
2
3
4
5
Figure 3.2
Removing Amplifier Unit