![YOKOGAWA AE100D Скачать руководство пользователя страница 241](http://html.mh-extra.com/html/yokogawa/ae100d/ae100d_instruction-manual_911083241.webp)
IM 1F6A0-01E
9-1
9. EXPLOSION PRTECTED TYPE INSTRUMENT
9. EXPLOSION PROTECTED TYPE INSTRUMENT
explosion proof type instrument are described except JIS
Flame proof. For explosion proof type instrument, the
description in this chapter is prior to other description in this
Instruction Manual.
WARNING
• Only trained persons use this instrument in
industrial locations.
9.1
CENELEC ATEX (KEMA)
WARNING
Only trained persons use this instrument in
industrial locations.
9.1.1
Technical Data
• Explosion proof
Type of protection: EEx d IIC T6…T1 (Integral Type
Flowmeter and Remote Type Flowmeter)
EEx d IIC T6 (Remote Type Converter)
Groups: Group II
Category: Category 2G
Temperature Code: (Integral Type Flowmeter and Remote
Type Flowmeter)
Ambient Temperature Process Temperature
+60
°
C
85
°
C
+60
°
C
100
°
C
+60
°
C
135
°
C
+60
°
C
200
°
C
+60
°
C
300
°
C
+60
°
C
450
°
C
T6
Temperature Class
T5
T4
T3
T2 *1
T1 *1
T090101-1.eps
*1
Note: Use /HT version above 260
°
C
Temperature Code: T6 (Remote Type Converter)
Drgree of Protection of Enclosure: IP67
Tamb: -29 to +60
°
C (Integral Type Flowmeter and
Remote Type Flowmeter)
-30 to +60
°
C (Remote Type Converter)
-29 to +60
°
C (Integral Type Flowmeter with indicator)
-30 to +60
°
C (Remote Type Converter with indicator)
Ambient Humidity: 0 to 100%RH
Power Supply: 42Vdc max.
Output Signal: Current Output; 4 to 20mAdc
Pulse output; On=2Vdc, 200mA
Off=42Vdc, 4mA
Maximim Working Pressure: 16MPa(DY015 to DY200)
5MPa(DY200 and DY300)
Coating of Enclosure: Epoxy resin coating or Polyurethane
resin coating
Electrical Connection: ANSI 1/2 NPT female,
ISO M20
X
1.5 female
• Intrinsically Safe
Type of protection: EEx ia IIC T4…T1 (Integral Type
Flowmeter and Remote Type Flowmeter)
EEx d IIC T4 (Remote Type Converter)
Groups: Group II
Category: Category 1G
Maximum Working Pressure: 16MPa (DY015 to DY200)
5MPa (DY250 and DY300)
Tamb. (Integral Type Flowmeter) : -29 to +60
°
C
Tamb. (Remote Type Flowmeter) : -29 to +80
°
C
Tamb. (Remote Type Converter) : -40 to +60
°
C
Ambient Humidity: 0 to 100%RH (No condensation)
(Integral Type Flowmeter)
Ambient Temperature Process Temperature
60
°
C
135
°
C
60
°
C
200
°
C
60
°
C
300
°
C
60
°
C
450
°
C
T4
Temperature Class
T3
T2*
T1*
T090101-2.eps
*: Use /HT version above 260
°
C
(Remote Type Flowmeter)
Ambient Temperature Process Temperature
80
°
C
135
°
C
80
°
C
200
°
C
80
°
C
300
°
C
80
°
C
450
°
C
T4
Temperature Class
T3
T2*
T1*
T090101-3.eps
*: Use /HT version above 260
°
C
For connection to certified Intrinsically Safe circuit with
Signal/Supply and Pulse circuit of DY(/HT) and DYA :
Ui=30V dc, Ii=165mAdc, Pi= 0.9W, Ci= 6nF, Li= 0.15mH
Connect sensor circuit of DYA to DY-N(/HT) :
Maximum cable capacitance : 160nF
Electrical Connection:ANSI 1/2 NPT female, ISO M20
X
1.5
female
In this section, further requirements and differences for
Содержание AE100D
Страница 2: ......
Страница 333: ...Blank Page ...
Страница 395: ...Blank Page ...
Страница 401: ...Blank Page ...