IM 1F6A0-01E
6-10
6. PARAMETERS
Output
(%)
Setting Value
5% Hysteresis
Contact Output ON : FLOW SW (LOW : ON)
Contact Output OFF : FLOW SW (LOW : OFF)
F060302.EPS
[B30:UPPER DISP] Upper indicator display
Select upper display, Flow rate (%) (0), Flowrate (1),
TEMP(%)(2). “TEMP(%)” can be selected when Option
Code
/MV.
[B31:LOWER DISP] Lower indicator display
Select lower indicator display, “BLANK (0), TOTAL (1),
TEMP(2). When “BLANK” in B31 is selected, indicator is
blank. “TEMP” can be selected when Option Code
/MV.
[B40:TOTAL START]
Select the START/STOP of totalizer from “STOP (0),
START (1).”
[B45:TOTAL RATE] Total rate of the totalizer
Set the total rate of the totalizer.
[B47:TOTAL RESET] Reset the totalizer
When totalizer reset function is excuted, the total display and
communication parameter are reset.
The following items should be done in case of
which Option code “/MV” is selected.
[B50 A/OUT SELECT] Analog Output select
Select the analog output select from flow rate or temperature.
When changing the analog output, UPPER DISPLAY can be
changed shown as below automatically.
B50 : A/OUT SELECT
“TEMP” TO “FLOW”
“FLOW” TO “TEMP”
UPPER DISPLAY
FLOW(%)
TEMP(%)
(“B30 : UPPER DISPLAY” is “FLOW RATE”, it can not be changed.)
T060401_1.EPS
The following item should be done in case of
which B50 is “TEMP”
[B51 TEMP 0%]
Set temperature value of 0% output.
[B52 TEMP 100%]
Set temperature valve of 100% output.