![YOKOGAWA AE100D Скачать руководство пользователя страница 26](http://html.mh-extra.com/html/yokogawa/ae100d/ae100d_instruction-manual_911083026.webp)
IM 1E7D0-01E
5-2
5. OUTLINE
Size in mm
(inch)
Span in m/s
(ft/s)
Accuracy
2.5 to 15
(0.1 to 0.5)
0.3 to 1
(1 to 3)
1 to 10
(3 to 33)
0.3 to 1
(1 to 3)
1 to 10
(3 to 33)
25 or larger
(1)
0.5% of span
0.25% of span
(at indications below 50% of span)
0.5% of flowrate
(at indications 50% of span or more)
0.25% of span
(at indications below 50% of span)
0.5% of flowrate
(at indications 50% of span or more)
0.1% of span
(at indications below 20% of span)
0.5% of flowrate
(at indications 20% of span or more)
T04.EPS
Polyurethane Lining
Size in mm
(inch)
Span in m/s
(ft/s)
Accuracy
0.3 to 1
(1 to 3)
1 to 10
(3 to 33)
25 or larger
(1)
0.5% of span
0.25% of span
(at indications below 50% of span)
0.5% of flowrate
(at indications 50% of span or more)
T05.EPS
Repeatability: 0.1% of flowrate (minimum 1 mm/s)
Maximum Power Consumption: 12.5W
Insulation Resistance:
• 100M
Ω
between excitation terminal and
each signal terminal at 500 V DC.
• 100M
Ω
among signal terminals at 500V
DC.
Withstand Voltage :
• 1500 V AC between excitation terminal
and ground terminal for 1minute. (for ex-
proof type).
• 500 V AC between each signal terminal
and ground terminal for 1minute.(for ex-
proof type).
• 1760 V AC between each signal terminal
and each excitation terminal for 1minute.
(for ex-proof type).
• 1000 V AC between excitation terminal
and ground terminal for 1minute.(except
for ex-proof type).
CAUTION
When performing the Voltage Breakdown Test,
Insulation Resistance Test or any unpowered
electrical test, wait 10 seconds after the power supply
is turned off before removing the housing cover. Be
sure to remove the Short Bar at terminal “G”. After
testing, return the Short Bar to its correct position.
Screw tightening torque should be 12kgf
-
cm(0.88ft-
lb)or more, because the G-terminal is thought as a
protective grounding and should conform to the Safety
Requirements.
Safety Requirement Standards:
IEC1010, EN61010
EMC Conformity Standards:
EN55011 Class A, Group 1
EN50082-2
AS/NZS 2064
■
NORMAL OPERATING CONDITIONS
Ambient Temperature: -20 to 60˚C (–4 to 140˚F)
Ambient Humidity: 5 to 95%RH (no condensation)
Power Supply Voltage: -A1; Range 80 to 264 V AC, 47
to 63Hz/100 to 130 VDC,
-D1; Range 20.4 to 28.8 V DC
Supplied Power and Max. Cable Length for 24 V DC
version:
Allowable cable length m(ft)
1000(3300)
900(2970)
800(2640)
700(2310)
600(1980)
500(1650)
400(1320)
300(990)
200(660)
100(330)
0
20
22
24
26
28
(V)
Cable cross section area: 1.25 mm
2
Cable cross section area: 2 mm
2
F01.EPS
Altitude at installation side: Max.2000m above sea
level
Installation category based on IEC1010: II(See Note)
Pollution level based on IEC1010: 2(See Note)
Note: • The “Installation category“ implies the regulation
for impulse withstand voltage. It is also called
the“Overvoltage category“. “II“ applies to electrical
equipment.
• “Pollution level“describes the degree to which a
solid, liquid or gas which deteriorates dielectric
strength is adhering. “2“ applies to normal indoor
atmosphere.
Fuse: 2A, 250V(Time - Lag type)
Fluid Conductivity:
Size 15 to 200mm(0.5 to 8in.); 5
µ
S/cm or larger
Size 2.5 to 10mm(0.1 to 0.4in.); 20
µ
S/cm or larger
Measurable Flow Rate Range:
SI Units (Size: mm, Flowrate: m
3
/h)