2. GENERAL DESCRIPTION
2.4 Option Specifications
T020401.EPS
Code
Specification
JIS Flame proof Approval (Note 1)
Flame proof Ex d IIC T6 Certified by TIIS.
(TIIS is the abbreviation of Technology Institution of Industrial Safety.)
Amb. Temp: –20 to 60
ⴗ
C
Electrical connection : JIS G1/2 female
FM Explosion proof Approval
Type of Protection : Explosionproof for Class I, Division 1, Groups A, B, C and D;
Dust-ignitionproof Class II/III, Division 1, Groups E, F, and G.
"SEAL ALL CONDUITS WITHIN 18 INCHES." "WHEN INSTALLED IN DIV.2,
SEALS NOT REQUIRED."
Enclosure Rating : NEMA TYPE 4X
Temperature Code : T6
Ambient Temperature : –29 to 60
ⴗ
C (Integral Type Flowmeter and Remote Type Flowmeter)
–40 to 60
ⴗ
C (Remote Type Converter)
Ambient Humidity : 0 to 100%RH
Maximum Working Pressure : 15MPa(2160 psi) (DY015 to DY200)
5MPa(720 psi) (DY250 and DY300)
Coating of Enclosure : Epoxy resin coating or Polyurethane resin coating.
Electrical Connection : ANSI 1/2NPT female
FM
Intrinsically safe Approval (Note 2)
Type of Protection : Intrinsically Safe for Class I, II, III, DIV.1, Groups A, B, C, D, E, F and G, T4,
and Class I, Zone 0, AEx ia IIC T4
Nonincendive for Class I, II, Div.2, Groups A, B, C, D, F and G, Class III, DIV.1, T4,
and Class I, Zone 2, Groups IIC, T4
Ambient Temperature : –29 to +60
ⴗ
C
(Integral Type Flowmeter)
–29 to +80
ⴗ
C (Remote Type Flowmeter)
–40 to +80
ⴗ
C (Remote Type Converter)
Ambient Humidity : 0 to 100% RH (No condensation)
Indoors and Outdoors : NEMA TYPE 4X
Electrical Parameters : Vmax=30Vdc Imax=165mAdc, Pi=0.9W, Ci=12nF, Li=0.15mH
Electrical Connection : ANSI 1/2NPT female
CENELEC ATEX(KEMA) Explosionproof Approval
Type of protection : EExd IIC T6...T1(Integral Type Flowmeter and Remote Type Flowmeter)
EExd IIC T6 (Remote Type Converter)
Groups : Group IIG
Category : Category 2
Temperature Class : T6...T1(Integral Type Flowmeter and Remote Type Flowmeter)
T6(Remote Type Converter)
Process temp.: T6; 85
ⴗ
C, T5;100
ⴗ
C; T4;135
ⴗ
C; T3;200
ⴗ
C;T2;300
ⴗ
C; T1;450
ⴗ
C
(Use /HT version above 260
ⴗ
C)
Degree of Protection of Enclosure : IP67
Tamb: –29 to +60
ⴗ
C (Integral Type Flowmeter and Remote Type Flowmeter)
–30 to +60
ⴗ
C (Remote Type Converter)
–29 to +60
ⴗ
C (Integral Type Flowmeter with indicator)
–30 to +60
ⴗ
C (Remote Type Converter with indicator)
Ambient Humidity : 0 to 100% RH
Maximum working Pressure : 16MPa(DY015 to DY200)
5MPa(DY250 and DY300)
Coating of Enclosure : Epoxy resin coating or Polyurethane resin coating.
Electrical Connection : ANSI 1/2 NPT female, ISO M20 X 1.5 female.
CENELEC ATEX(KEMA) Intrinsically safe Approval (Note 2)
Type of protection : EEx ia IIC T4...T1(Integral Type Flowmeter and Remote Type Flowmeter)
EEx ia IIC T4(Remote Type Converter)
Groups : IIG
Category : 1
Maximum Working Pressure : 16MPa (DY015 to DY200)
5MPa (DY250 and DY300)
Tamb.(Integral Type Flowmeter) : –29 to +60
ⴗ
C
Tamb.(Remote Type Flowmeter) : –29 to +80
ⴗ
C
Tamb.(Remote Type Converter) : –40 to +60
ⴗ
C
Ambient Humidity : 0 to 100%RH (No condensation)
Process temp.: T4;135
ⴗ
C; T3;200
ⴗ
C; T2;300
ⴗ
C; T1; 450
ⴗ
C
(Use /HT version above 260
ⴗ
C)
For connection to certified Intrinsically Safe circuit with Signal/Supply and Pulse circuit of
Integral Type Flowmeter and Remote Type Converter : Ui=30Vdc, Ii=165mAdc, Pi=0.9W, Ci=6nF,
Li=0.15mH
Connect sensor circuit of DYA and DY-N(/HT) : Maximum cable capacitance:160nF
Electrical connection : ANSI 1/2NPT female, ISO M20 X 1.5 female.
Item
Japanese Industrial
Standards (JIS)
Factory Mutual (FM)
CENELEC ATEX
(KEMA)
(Note 1) In case of JIS Flameproof (/JF3), specify in the option code with Flame proof packing ground (/G11, /G12) for the cable wire
construction. In case the ambient temperature exceeds 50deg.C, use heat resistant cables with maximum allowable
temperature of 70degC or above.
(Note 2) For intrinsically safe approval, use the barrier certified by the testing laboratories (BARD-400 is not applicable).
JF3