WIKA operating instructions pneumatic high pressure controller, model CPC7000
0019104001D 10/2018EN
EN
Localgravity?
<sp>+n.nn<cr><lf>
Returns the local gravity in feet/sec^2
LowerLimit
Value inside primary xducer range in current units.
Sets the lower control limit for the instrument
LowerLimit?
<sp>n.nn<cr><lf>
Returns the lower control limit for the instrument in
current units.
Macaddress?
<sp>cc-cc-cc-cc-cc-cc<cr><lf>
Returns Ethernet MAC address
Measure
None
Instrument placed in Measure Mode
Measure?
<sp>(YES or NO)<cr><lf>
Returns YES if instrument is in measure. NO if other-
wise
Mode
MEASURE, CONTROL, VENT, STANDBY
Sets the operation mode
Mode?
<sp>XXXXXX<cr><lf>
Returns the operation mode
Netmask
nnn.nnn.nnn.nnn
Sets the Ethernet network mask
Netmask?
<sp>nnn.nnn.nnn.nnn<cr><lf>
Gets the Ethernet network mask
Outform
1 to 7 – see table in Section 7.10
Sets the output format
Outform?
<sp>X<cr><lf>
Returns the output format – see table in Section 7.10
Peakmax?
<sp>n.nn<cr><lf>
Returns the maximum pressure since peakreset was
sent
Peakmin?
<sp>n.nn<cr><lf>
Returns the minimum pressure since peakreset was
sent
Peakreset
None
Resets the peak values.
Port
nnnnnn
Sets the Ethernet port of the instrument
Port?
<sp>nnnnn<cr><lf>
Returns the Ethernet port of the instrument
Ptype
Absolute or Gauge
Sets the instrument pressure type – emulation only
works if the optional barometric transducer is installed.
If no barometer is present, standard atmosphere is
used for the barometric reading when in emulation
mode.
Ptype?
<sp>CCCCC<cr><lf>
Returns “Absolute” or “Gauge” for the pressure type
in native mode and “Absolute Emulated” or “Gauge
Emulated” for the emulated pressure types
RangeMax?
<sp>n.nn<cr><lf>
Returns the maximum range of the active transducer in
the current units.
RangeMin?
<sp>n.nn<cr><lf>
Returns the minimum range of the active transducer in
the current units.
Rate?
<sp>n.nn<cr><lf>
Returns the rate reading of the instrument in current
units/current time unit (see Runits)
Rdecpt?
<sp>n<cr><lf>
Returns the number of rate decimal points. (see
Resolution)
Release?
15 char string.
EX:
Release? Test_stand_1
Returns:
<sp>(YES or NO), CCC…CCC<cr><lf>
This command is used to release control of the instru-
ment in a multiple computer environment.
Yes if release is successful
No if instrument is being controlled with another
computer
CCC… = name of controlling computer or AVAILABLE
(see Acquire? and Unlock)
Resolution
<n>
Sets the number of significant digits. (see decpt)
Resolution?
<sp>n<cr><lf>
Returns the number of significant digits. (see decpt)
Rfilter
Value in %
Sets the % of the rate filter
Rfilter?
<sp>n.nn<cr><lf>
Returns the rate filter.
Rsetpt
Value in current units
Sets the rate setpoint
Rsetpt?
<sp>n.nn<cr><lf>
Returns the rate setpoint
7. Remote Operation