WIKA operating instructions pneumatic high pressure controller, model CPC7000
0019104001D 10/2018 EN
EN
5. Installation
5. Installation
Personnel
: Skilled electrical personnel
WARNING!
READ THSE INSTRUCTIONS BEFORE INSTALLATION!
Only use original parts (see chapter 11 “Accessories”).
CAUTION
Damage to the instrument
When working on open electrical circuits (printed circuit boards) there is a risk of damaging sensitive electronic
components through electrostatic discharge.
▶
The correct use of grounded working surfaces and personal armbands is required.
5.1 Mounting
The instrument can be set up on a desk top or it can be rack-mounted. Rack mount hardware is optional on the CPC7000 (see
Section Specifications11 and Section 12 Options and Accessories).
The special transducers used in the CPC7000 are relatively insensitive to tilt and vibration. However to further assure stability and
accuracy, avoid mounting the instrument on surfaces subject to excessive motor or machinery vibration
5.2 Rear Panel
Up to five pneumatic pressure ports are located across the bottom of the rear panel. The instrument with gauge transducers will
have a reference and a pressure port. The instrument with absolute transducers will use the pressure port. In the upper left corner
under the digital I/O connector is a M12 fitting , which is connected to the barometric reference sensor if installed. The connections
for digital I/O - 1 & 2 are located on the top on left side of the ventilation fan. Positioned on the right side is the RS-232, Ethernet,
IEEE- 488, USB device connections for communication, as well as a USB host connection and the 12 VDC power input.
Figure 5.2 Rear Panel
5.2.1 Rear Panel
WARNING!
The pressure connections must be installed according to the following instructions, observing the relevant
regulations. The installation is to be performed by trained, authorized personnel, knowledgeable in the safety
regulations for working on pneumatic/hydraulic systems
Up to 7 pressure connections are on the rear panel. Pressure connections that are not assigned are plugged.