WIKA operating instructions pneumatic high pressure controller, model CPC7000
0019104001D 10/2018EN
EN
11. Specifications
Accuracy specifications presented herein are obtained by comparison with primary standards traceable
to a national metrology institute or recognized international standard organization. These specifications
are obtained in accordance with the ISO Guide to the Expression of Uncertainty in Measurement (GUM).
The calibration program at Mensor is accredited by the American Association of Laboratory Accreditation
(A2LA) as complying with both the ISO/IEC 17025:2005 and the ANSI/NCSL Z540-1-1994 standards. If
there is an exception to the requirements and recommendations of Z540 during a calibration the exception
is noted on the individual calibration certificate.
Mensor reserves the right to change specifications without notice.
11.1 Measure Specification
Reference pressure transducers
Pressure range
CPR8000
CPR8050
Accuracy
1)
Standard: 0.01 % FS
2)
Option: 0.01 % IS-50
3)
0.01 % FS
2)
Gauge pressure
100 ... 400 bar
(1,500 ... 6,000 psi)
4)
400 ... 700 bar
(6,000 ... 10,000 psi)
4)
Absolute pressure
101 ... 401 bar abs.
(1,515 ... 6,015 psi abs.)
4)
401 ... 701 bar abs.
(6,015 ... 10,015 psi abs.)
4)
Calibration interval
365 days
365 days
Optional barometric reference sensor
Function
The barometric reference sensor can be used to switch pressure types
5)
, absolute <=> gauge.
Measuring range
552 ... 1,172 mbar abs. (8 ... 17 psi abs.)
Accuracy
1)
0.01 % of reading
Pressure units
39 and two freely programmable
1)
It is defined by the total measurement uncertainty, with the coverage factor (k = 2) and includes the intrinsic performance of the instrument, the measurement uncertainty of the reference
instrument, long-term stability, influence of ambient conditions, drift and temperature effects over the compensated range with recommended zero point adjustment every 30 days.
2) FS = full span
3) 0.01 % IS-50 accuracy: Between 0 ... 50 % of the full scale, the accuracy is 0.01% of half of the full scale value and between 50 ... 100 % of the full scale, the accuracy is 0.01 % of reading.
4) Ranges from 1500 to 2000 psig will be sealed gauge transducers.
5) For a pressure type emulation, we recommend a native absolute pressure transducer, since the zero point drift can be eliminated through a zero point adjustment.
11.2 Base Instrument
Base Instrument
Instrument
Instrument version
Standard: desktop
Option: 19” rack-mounting kit
Dimensions
See technical drawings
Weight
approx. 40 kg (88.2 lbs) incl. all internal options
Warm-up time
approx. 15 min
Display
Screen
8.9” color LCD with resistive touchscreen
Resolution
4 ... 6 digits depending on range and units
Connections
Pressure connections
Standard: 5 ports with M16 x 1.5 female with sealing cone
Optional: 2 ports with 9/16-18 UNF female
Optional barometer: M12 x 1 female thread
Filter elements
The instrument has a 20-micron filters on all pressure ports.
11. Specifications