60
WIKA operating instructions pneumatic high pressure controller, model CPC7000
0019104001D 10/2018EN
EN
Burst_Slow
Value inside upper and lower limits
This is the rate used during the transition from low
setpt to high setpt in the anticipated “burst range”
Burst_Slow?
<sp>n.nn<cr><lf>
Returns the slow rate for burst test
Burst_Fast
Value inside upper and lower limits
This is the rate used during the transition from atmo-
sphere to the low setpt
Burst_Fast?
<sp>n.nn<cr><lf>
Returns the fast rate for burst test
Burst_Result?
<sp>n.nn<cr><lf>
Pressure at which burst was detected. Will return 0 if
no burst found.
Calculate_as_found_linearity
Calculate linearity slopes and intercepts from true/
actual pressures
Caldisable
YES,NO
Sets whether or not calibration of the active transducer
is disabled.
Caldisable?
<sp>(YES or NO)<cr><lf>
Returns whether or not calibration of the active trans-
ducer is disabled.
Cerr
None
Clears the error queue
Cmdset
Mensor, SCPI
Activates remote command set for instrument emula-
tion modes.
Cmdset?
<sp><CCCCCC><cr><lf>
Returns active command set identifier
Control
Instrument placed in Control Mode
Control?
<sp>(YES or NO)<cr><lf>
Returns YES if instrument is in control. NO if otherwise
Crate
Slow, Medium, Fast, Variable
Sets the control rate, Variable mode is a predeter-
mined user defined rate setpt.
Crate?
<sp>CCCCCC<cr><lf>
Returns the control rate – CCCC is variable in length
and corresponds to the parameters for the CRATE
command
Ctype?
Returns the type of regulator
Decpt?
<sp>n<cr><lf>
Returns the number of decimal points (see Resolution)
Default
None
Sets the default values
DHCP
ON or OFF
Reserved for DHCP setup
DHCP?
<sp>(YES or NO)<cr><lf>
Reserved for DHCP setup
DIO
2 or 0
2 turns on the first digital output pin, 0 turns it off
DIO?
<sp>n<cr><lf>
Returns status of the first input and output pin. Bit0 =
input’s status, Bit1 = output’s status.
DOUTFUNC
<n><sp><CCCCCC><cr><lf>
Sets the function for output pin <n> to NONE, STA-
BLE, MEASURE, CONTROL, VENT
DOUTFUNC? <n>
<sp><CCCCCC><cr><lf>
Returns the function for output pin <n>
DINFUNC
<n><sp><CCCCCC>
Sets the function for input pin <n> to NONE, MEA-
SURE, CONTROL, VENT, KEYLOCK, START,
READING
DINFUNC? <n>
<sp><CCCCCC><cr><lf>
Returns the function for output pin <n>
DOUTSTATE
<n><sp><HIGH/LOW/1/0>
Turns output pin <n> to high, or low. Sets the function
for that pin to “None”
DIOSTATE?
<sp><n><cr><lf>
Bit 0-5 is the status of the Input bits, Bit 6-11 are the
status of the Output bits. Returns an integer between
0 and 4095
DOC
mm/dd/yyyy
Sets the date of cal for the active transducer.
DOC?
<sp>mm/dd/yyyy<cr><lf>
Returns the date of cal for the active transducer.
DOM?
<sp>mm/dd/yyyy<cr><lf>
Returns the date of manufacture
Error?
<sp> text description <cr><lf>
Returns the next error in the error queue
7. Remote Operation