38
WIKA operating instructions pneumatic high pressure controller, model CPC7000
0019104001D 10/2018EN
EN
6.4.3.6
Vent Limit
This is the pressure at which controlled vent stops and the vent solenoid is opened causing the pressure to vent through the vent
NO port uncontrolled. The vent limit can be set within the values of the primary transducer.
Figure 6.4.3.6 Vent Limit
6.4.3.7
Control Volume
The Control Volume button in the Control Settings App allows the user to set the control pressure volume in cubic centimeters (cc)
between 0 ... 50 cc. By default this is set to 25 cc. Figure 6.4.3.7 displays the control volume setting for the instrument.
Figure 6.4.3.7 Control Volume
6. Operation