17
WIKA operating instructions pneumatic high pressure controller, model CPC7000
0019104001D 10/2018 EN
EN
4. Transport, Packaging and Storage
EU).
3.8 Warnings and Cautions
WARNING!
HIGH PRESSURE!
High pressure gases are potentially hazardous. Energy stored in these gases and liquids can
be released suddenly and with extreme force. High pressure systems should be assembled and operated only by
personnel who have been trained in proper safety practices.
WARNING!
NOT EXPLOSION PROOF!
Installation of this instrument in an area requiring devices rated as intrinsically safe is not
recommended.
WARNING!
POSSIBLE INJURY!
The tubing, valves, and other apparatus attached to the gauge must be adequate for the
maximum pressure which will be applied, otherwise physical injury to the operator or bystanders is possible.
CAUTION
USE THE PROPER PRESSURE MEDIUM!
Use only clean, dry, nitrogren of class 2.8 or better unless otherwise
specified by Mensor. This instrument is not designed for air use.
CAUTION
As with most sensitive electronic equipment, switch the power switch off before connecting or disconnecting to a
power source to prevent data loss. Do not position the equipment so that it is difficult to disconnect the AC power
cord.
WARNING!
Detachable main power supply cord delivered with inadequate ratings should not be used, it is recommended to use
the power cord delivered with the instrument. See Section 11 Specifications for power ratings.
Additional Warning and Caution notices are found throughout this manual.