56
WIKA operating instructions pneumatic high pressure controller, model CPC7000
0019104001D 10/2018EN
EN
7. Remote Operation
7.2 Remote Operating Parameters
Use the screens in Section 6.4.5 Remote Application to set the remote operating parameters for the instrument command set,
Ethernet, Serial (RS-232) and IEEE-488 (GPIB) communications.
7.3 Command Set
Command Set button – Users’ can select which model remote protocol they would like to emulate for simulation and testing
purposes. Selections may include the following or may be added per customers’ specifications:
■
Mensor (default)
■
SCPI WIKA (The SCPI WIKA mode emulates the WIKA command set in SCPI format.)
7.4 IEEE-488
IEEE-488 address button – Allows the user to set the GPIB address by inputting a numeric value utilizing the touch screen.
7.4.1 IEEE-488.2 Commands
Command
Response/ Function
*IDN?
Returns identification string
7.5 Ethernet
The Ethernet function allows the user to set the following by inputting a numeric value in each separate field:
■
IP
■
Netmask
■
Gateway
■
Port
■
DHCP settings
Set the Ethernet communication parameters as described in Section 6.4.5.
CAUTION! Please contact your network administrator for
proper settings
.
CAUTION! Please consult your computer resources department prior to connecting this instrument to your
network to verify there are no conflicts with existing IP addresses
.
The Ethernet communication port allows the CPC7000 to communicate with computers using 10/100Based-T specifications.
Before using Ethernet communication, four parameters must be set up: IP, Netmask, Gateway, and Port.