80
WIKA operating instructions pneumatic high pressure controller, model CPC7000
0019104001D 10/2018EN
EN
Figure 9.2.9-B shows some typical values that might be seen in a linearization calibration. In the bottom right hand corner of this
screen is the Graph Icon [
] that, when pressed, reveals a Linearization Error Graph (figure 9.2.9-C) that gives a visual
representation of the errors associated with the values entered in the Linearization screen.
This Linearization error graph shows a scaling that corresponds to the maximum error calculated from the data entered in the
Linearization Matrix. It is a good indication of the overall error of the transducer, and will quickly reveal any gross data entry errors
that have been made. To revert back to the Linearization Matrix press the Matrix Icon [
].
Figure 9.2.9-C Linearization Values
Note: After calibration is complete, return to the Calibration Data Application (Section 9.6) to record the certificate
number, calibration interval and the date of calibration. Restoration to factory calibration can also be completed in
this application
9.2.10 Head Pressure
The Head Pressure Application provides an automated way to calculate the head pressure offset between a device
being tested and the CPC7000 transducer, based on
■
Media: pneumatic
■
Height: the difference between the device under test and the transducers in the CPC7000
(Difference = CPC7000-DUT)
■
Gas Density: the density of the pressure media
■
Gas Temperature
■
Local Gravity
The Head Pressure application is placed under the password protected area of the operator interface to safeguard against
inadvertent activation. It is intended to be used to accurately measure the pressure at an external source that is at a different
elevation from the CPC7000 transducer sensing that pressure.
The four parameters are used to calculate the pressure that is a result of the different elevations. It should not be used when
calibrating CPC7000 transducers. The Head height should be set at zero before calibrating the transducers of the CPC7000.
9. Maintenance and Recalibration