6.4.4.2
Reading Resolution
The Resolution of the displayed pressure value can be set in the Display Settings Application using the resolution Parameter. The
resolution can be set to 4, 5 or 6 digits.
Figure 6.4.4.2 Reading Resolution
6.4.4.3
Bar Graph
The Bar Graph visible on under the displayed pressure value can be turned on/off in the Display Settings Applicaiton. The bar
graph indicates the relative position of the current value with the maximum range of the primary transducer.
6.4.4.4
Cal Function
The Cal Function presents a choice of None, Tare or Zero. Choosing Zero will enable the Zero Cal Button [
] in the Home
App. Choosing Tare will enable the Tare Button [
] in the Home App. The Tare button and the Zero Button cannot appear
on the screen at the same time. See Section 6.4.1.6 and 7.4.1.7 for operation of the Zero and Tare buttons in the Home App (main
screen).
Figure 6.4.4 Cal Function