Troubleshooting
-80-
7-2 Troubleshooting
Problem Cause Remedy
The AC adapter is not connected.
- Check that the AC adapter is
connected.
- If nothing is displayed even if
the AC adapter is properly
connected, the electric
components of the balance, or
the AC adapter may have
failed. Replace the AC adapter
to test the root cause of the
problem, if you have a
same-model AC adapter that
operates properly at hand.
- Contact our Sales Office or
Technical Service Division.
Nothing is displayed even when
the balance is powered on.
The batteries are exhausted.
Replace the batteries.
Display flickers.
The balance may be affected by
an external influence such as
wind and vibration.
Increase the setting values of
relevant functions by referring to
“Section 5-6: Improving the
Stability of the Balance” (P. 59).
The display error is caused
because the balance has not been
used for a long period of time or
has been relocated to another
location.
Perform span adjustment.
The adjusters are not settled, and
the balance is not kept
horizontal.
Check that the balance is kept
horizontal.
Weight indication contains an
error.
The tare weight is set or not.
Unload the sample from the pan
and then zero the readout by
pressing the [Zero/Tare] key to
continue measurement.
Содержание CJ-E series
Страница 2: ......
Страница 9: ...Safety Precautions vii Dispose of batteries in accordance with local regulations...
Страница 14: ......
Страница 24: ......
Страница 32: ......
Страница 60: ......
Страница 74: ...Function Setting 60 4 Save the setting Press the Set key The balance goes back to normal display...
Страница 90: ......