3-5 Measuring Specific Gravity
-31-
Using a liquid medium other than water
1
Set to the function setting mode.
(Refer to “Section 2-5: Function Setting Basics (P.
16).)
Press and hold the [Function] key. After “
”
is displayed, release the finger.
2
Select “Gravimeter.”
Press the [Zero/Tare] key several times to select
“
.”
3
Set to the medium select mode.
Press the [Function] key.
“
” is displayed.
4
Select “Not water.”
Select “1” (Not water) by pressing the [Zero/Tare]
key and then press the [Set] key.
Now, the specific gravity in a medium other than water
can be measured.
Measure specific gravity using the steps described in
“Measurement” (P. 29). In step 5, set the density of the
used liquid medium instead of water temperature. - The
acceptable range of liquid density is 0.0001 to 9.999
g/cm
3
.
Gravimeter measurement data output
Measured specific gravity data is output to a printer as follows depending on settings:
- Before
measurement
Irrespective of the setting made in the function setting “
” (output control), pressing
the [Print] key outputs data (irrespective of whether data is stable or unstable).
- While specific gravity is displayed
Output contents and methods can be changed by the function setting “
” (output
data select) and “
” (auto output).
- Output
format
If [1] (specific gravity, weight, and actual water temperature or density of media) is set in the
function setting “
” (output data select), all data is output (See the figures below).
If [0] (specific gravity only) is set in the function setting “
” (output data select),
only the first and second line values shown in the figures below are output.
If any statistical calculation is made for the printer, the values in the second line shown in the
figures below will be printed being prefixed by a serial number.
Содержание CJ-E series
Страница 2: ......
Страница 9: ...Safety Precautions vii Dispose of batteries in accordance with local regulations...
Страница 14: ......
Страница 24: ......
Страница 32: ......
Страница 60: ......
Страница 74: ...Function Setting 60 4 Save the setting Press the Set key The balance goes back to normal display...
Страница 90: ......