7-3 Initializing
-83-
7-3 Initializing
The settings of the balance can be initialized with the following steps:
1
Set to the function 2 setting mode.
+
Press the [Function] key while pressing the
[Zero/Tare] key. Release the fingers when “
” is displayed.
The function 2 setting mode is set, displaying “
.”
2
Set the function initialization.
Select “
” by pressing the [Function] key.
Select “1” by pressing the [Zero/Tare] key.
3
Save the setting.
Press the [Set] key.
The balance goes back to normal display.
注 意
All function settings are initialized, discarding all data including ID numbers, limit
values, data of parts counting, percentage weighing, unit converting, and gravimeter.
Restoring the current status will be impossible. Before initializing the balance, record
necessary function settings in a memo or otherwise.
Caution